Company record
C-RAD POSITIONING AB
0 FDA warning letters since 2023 and 1 recall in the last 12 months.
Get an email the day FDA posts anything new about C-RAD POSITIONING AB.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jan 21 | Class II | C-RAD POSITIONING AB Device |
Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioni… Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. FDA notice Z-1092-2026 |